GE DRAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
Chapter 3 - Setup Instructions 3-75
Section 3-13Network Configuration
3-13-1 TCP/IP Configuration
NOTE: Following Information must be provided by customer or hospital engineer before you can start:
A Station name, AE Title, IP address and Port Number for the Voluson™ P8 / Voluson™ P6.
The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION.
Only if necessary (e.g. for Internet access).
1.) Press the UTILITIES
key on the control console.
2.) In the “Utilities” menu select SETUP
button to invoke the setup desktop on the screen.
3.) On the left side of the screen select CONNECTIVITY
and then click the DEVICE SETUP tab.
4.) Click the button, read the message and confirm with YES
.
5.) The “Internet Protocol (TCP/IP) Properties” dialog page appears.
Figure 3-76 Network Configuration