Page #338 background imageLoading...
Page #338 background image
GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
7-46 Section 7-10 - Troubleshooting Trees, Instructions and Tech Tips
7-10-3 Trackball - Impaired Sensitivity
Figure 7-40 Trackball - Impaired Sensitivity
The trackball has low sensitivity!
NO
YES
Is it difficult to roll the
trackball?
Remove the trackball and clean it from dust.
Remove the trackball and clean it from dust.
Mount the outer fixing ring by turning it clockwise.
Close
Return to normal scanning
Contact GE service representative
to replace the trackball module.
Set up the appropriate trackball
response time
(via [Trackball Speed] button in the
System Setup - User Settings page).
Still low sensitive...
Fixing
Ring
Open