GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
5-50 Section 5-14 - Service Page
Section 5-14
Service Page
5-14-1 Introduction
The Service Page contains specific software/hardware test modules, system setup, update, etc. for
Voluson systems only.
5-14-2 Access / Security
The service page has different access and security user levels.
Each user is only granted access to the tools that are authorized for their use.
5-14-3 Service Login
1.) Press the UTILITIES key on the control console.
2.) In the “Utilities” menu select SETUP
button to invoke the setup desktop on the screen.
3.) Connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system.
As soon as the dongle is detected the login window pops up.Remove the dongle.
4.) Enter the “Standard GE Revolving” password (changes every six months) and then click ENABLE.
5.) On the left side of the screen select ADMINISTRATION and then click the SERVICE tab.
6.) Click the SERVICE TOOLS
button to display the “Service Tools” page.
Figure 5-25 Service Tools page
1. Common Service Desktop 7. Show Factory presets 13. Load Application Settings and Factory Presets
2. Auto Tester 8. Monitor Test 14. Keyboard Layout
3. Update 9. Hardware Test 15. Console test
4. Request for Service 10. Connected probes 16. Exit
5. Export System Data 11. Probe EEPROM
6. TCP/IP Buffersize 12. Delete all patients