GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
8-124 Section 8-39 - Replacement of the REAR HANDLE
Section 8-39
Replacement of the REAR HANDLE
8-39-1 Manpower
One person, 5 minutes
8-39-2 Tools
Philips screwdriver, Allen Wrench
8-39-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-39-4 Removal Procedure
1.) Remove the ASSY BEZEL FRONT PRINTER PLASTIC BT16. Refer to the figure below.
Figure 8-180 Removing the ASSY BEZEL FRONT PRINTER PLASTIC BT16
ASSY BEZEL FRONT PRINTER
PLASTIC BT16