8-12 Section 8-2 - System Software - Installation/Upgrade Procedure
21.)If Control Panel is not working after first boot up, please shut down the system; then boot up again.
22.)Reconnect the external devices, install all the printers and adjust the printer settings as described
in Section 3-7 "Printer Installation" on page 3-32.
23.)Check and match Printer Remote Control selection in the System Setup - Connectivity - BUTTON
CONFIGURATION page.
24.)Confirm that the date and time are set correctly and that the Windows automatic DST feature is ON.
25.)Perform Software and Functional checks as described in Section 8-3 on page 8-20.
8-2-3 Software DVD Storage
1.) Disassemble the Top Cabinet Cover.
2.) Put Software DVD in the jacket.
3.) Re-assemble the Top Cabinet Cover.
!! NOTICE:
After turning off a system, wait at least 10 seconds before turning it on again.
The system may not be able to boot if power is recycled too quickly.
Figure 8-11 Disassembling the Top Cabinet Cover
Figure 8-12 Putting Software DVD