GE DRAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-133
8-42-6 Serial Number - Reprogamming Procedure
Before calling OLC please prepare the following data:
• SNO of Voluson™ P8 / Voluson™ P6 (Alphanumeric serial number can be found on main label
at back of system)
• Hardware ID number (See SYSTEM INFO page or label of DRFM board)
1.) Gather System Serial number and Hardware ID
- Press the F2
key on upper OPIO, select SYSTEM SETUP, select ADMINISTRATION, click
on SYSTEM INFO and write down the "Hardware ID" number.
!! NOTICE:
To reprogramming the System Serial Number, contact the OLC(On Line Center) to get the necessary
"Serial Key" and "Permanent Key" (calculated in OKOS) !!!
Figure 8-193 System Setup - Administrator - System Info page
!! NOTICE:
The Hardware ID can also be found on the label on the DRFM board.