4-16 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-3 Save Full System Configuration (Full Backup)
A backup of the Full System Configuration always contains the following data
User Settings (databases and files containing User Programs, 2D/3D/4D Programs, Auto Text
entries, gray curves and complete System settings such as language, time/date format, etc.)
System Settings ( general “Setup” settings such as Language, Time/Date format, Button
configuration, Annotation settings, Biopsy lines, Peripheral data, Video Norm, Archive
configuration, etc.)
Measure Configuration (user specific measure setup settings)
Patient Archive (database containing patient demographic data and measurements) - no images
Options (Permanent Key that is specific for enabled software options and Demo Key)
Image Transfer Configuration (DICOM settings e.g., DICOM servers, AE Title, Station Name, etc.)
Network Configuration (network settings including the computer name)
Service Platform (state of the Service Software)
NOTE: It is recommended to “Full Backup” system configuration data before upgrading the software and/or
image settings (presets). This ensures that if settings need to be reloaded, will be the same ones the
customer was using prior to service.
1.) Press the F2
key on the control panel.
2.) In the “Utilities” menu select SYSTEM SETUP
to invoke the setup desktop on the screen.
3.) On the left side of the screen select BACKUP
and then click the SYSTEM CONFIGURATION tab.
4.) Click the SAVE
button of the “Full System Configuration” group.
5.) Choose the Destination.
6.) Enter the description of the full backup.
7.) Select the NEXT
button.
8.) To start the backup process click YES
.
After copying the data, the Voluson™ P8 / Voluson P6 reboots and the application starts again.
Figure 4-15 System Configuration Save
CAUTION
!! CAUTION:
Image data will not be backed up!
To backup the Image Archive, refer to Section 4-5-6 on page 4-21.
choose destination
enter description
click [Next] button

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P6 BT16 and is the answer not in the manual?

GE Voluson P6 BT16 Specifications

General IconGeneral
BrandGE
ModelVoluson P6 BT16
CategoryMedical Equipment
LanguageEnglish

Related product manuals