EasyManuals Logo

Druck DPI620G Series User Manual

Druck DPI620G Series
211 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #191 background imageLoading...
Page #191 background image
Copyright 2013 Baker Hughes Company.
English–DPI620G Instruction Manual | 181
Procedures (CH1/CH2): Current (Source)
a. Select the applicable channel Current (measure) function via the Calibrator Task
menu.
b. Apply these values:
mA: -
5
5, -25, -20, -10, -5, 0 (open circuit).
mA: 0, 5, 10, 20, 25, 55.
12.3 Procedures (CH1/CH2): Current (Source)
Follow this procedure:
1. Connect the applicable calibration equipment. Refer to Table 12-1.
2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on).
3. CH1 has only one range (24mA) while CH2 has two ranges (24mA and -24mA).
4. Use the Calibration menu (see Section 12.1) to do a two-point current source calibration
(0.2mA and FS) for the relevant channel:
CH1 (one range): 24mA
CH2 (two ranges): 24mA (reverse) and 24mA (forward).
Note: Input positive values for the forward and reverse calibration.
Figure 12-3: Calibration – Current Source (Range: +24mA and -24mA)
5. To make sure the calibration is correct:
a. Select the applicable Current (source) function via the Calibrator Task menu.
b. Apply these values: mA: 0.2, 6, 12, 18, 24.
Table 12-
2
: Current (Measure) Error Limits
Applied (mA) Calibrator Uncertainty (mA) Permitted DPI620G Error (mA)
± 55 0.00300 0.0055
± 25 0.00250 0.0040
± 20 0.00063 0.0022
± 10 0.00036 0.0016
± 5 0.00025 0.0013
0 (open circuit) 0.00020 0.0010

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Druck DPI620G Series and is the answer not in the manual?

Druck DPI620G Series Specifications

General IconGeneral
BrandDruck
ModelDPI620G Series
CategoryTest Equipment
LanguageEnglish

Related product manuals