Copyright 2013 Baker Hughes Company.
English–DPI620G Instruction Manual | 81
Set the Pressure Module to Zero
Figure 5-16: Example: Leak Test Results
Note: To set the Leak Test options, a pressure module or external pressure sensor must be
correctly installed.
5.8 Set the Pressure Module to Zero
SETTINGS > ZERO > ZERO
Use this option to write a new zero pressure value to the pressure module in use. The sensor zero
ad
justment is permitted only if the adjustment is less than 10% FS positive pressure value of the
sensor.
Note: To make a temporary adjustment for zero, use the Tare function.
Figure 5-17: Pressure Module Zero Example