GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL
2127661
RENEWAL PARTS
6–82
REV 9
6–2–8 Speaker Assy (FRU No. 112)
Time Required
0.5 Hours
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–24 on page 6–83.
1. Turn OFF the system.
2. Remove the Monitor Cover L Assy (FRU 104) and Monitor Cover R Assy (FRU 105). Refer to 6–2–2 on page
6–72.
3. Cut off the tie wrap fixing the CRT Cable Set (FRU 116) to the speaker.
4. Unscrew two screws (
1
and
2
).
5. Loosen two screws (
3
and
4
).
6. Disconnect one connector (
5
) from the back side of monitor.
7. Remove the speaker at the left.
8. Unscrew two screws (
6
and
7
).
9. Loosen two screws (
8
and
9
).
10. Disconnect one connector (
10
) from the back side of monitor.
11. Remove the speaker at the right.