2127661
LOGIQ 400 SERVICE MANUALGE MEDICAL SYSTEMS
RENEWAL PARTS
6–227
REV 9
6–2–92 Filter Cover Set (FRU No. 709)
Time Required
10 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–127.
1. Turn OFF the system and unplug the main power cord.
2. Remove the PW Air Filter.
3. Unscrew four screws (
1
4
).
4. Remove the Guard.
Note
The Filter Cover Set includes Retainer, PW Air Filter, and Guard.
FILTER COVER SET DISASSEMBLY
ILLUSTRATION 6–127