LOGIQ 400 SERVICE MANUALGE MEDICAL SYSTEMS
2127661
INSTALLATION
2–3
REV 0
2–1 PREINSTALLATION
2–1–1 Introduction
This section describes various general electrical, operational, and environmental considerations that must be consid-
ered before installing the LOGIQ 400 Ultrasound unit.
2–1–2 Power Line Requirements
The following power line parameters should be monitored for one week before installation. We recommend that you
use an analyzer Dranetz Model 606–3 or Dranetz Model 626:
PARAMETER : LIMITS
Voltage Range : Japan. : 100 VAC ±10% (90–110 VAC)
: Europe : 220–240 VAC ±10% (198–264 VAC)
USA : 120 VAC ±10% (108–132 VAC)
Power : Japan : MAX. 1000 VA
: Europe : MAX. 1000 VA
: USA : MAX. 1000VA
Line Frequency : All applications : 50/60Hz (±2Hz)
Power Transients : Less than 25 % of nominal peak voltage for less than 1 millisecond for any type of
transient, including line frequency, synchronous, asynchronous, or aperiodic
transients.
Decaying Oscillation : Less than 15 % of peak voltage for less than 1 millisecond.

Table of Contents

Other manuals for GE LOGIQ 400

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 400 and is the answer not in the manual?

GE LOGIQ 400 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 400
CategoryMedical Equipment
LanguageEnglish

Related product manuals