LOGIQ 400 SERVICE MANUALGE MEDICAL SYSTEMS
2127661
INSTALLATION
2–15
REV 0
2–2–10 Moving into Position
CAUTION
Do not lift the unit by the Keyboard.
Do not tilt the unit more than 5 degrees to avoid tipping it over.
To avoid injury by tipping over. Set the monitor to the lowest position before moving.
In general, a single adult can move the LOGIQ 400 along an even surface with no steep grades. At least two people
should move the machine when large humps, grooves, or grades will be encountered. (It is better to pull from the rear
rather than push from the front of the unit). Before moving, store all loose parts in the unit. Wrap transducers in soft
cloth or foam to prevent damage.
Although LOGIQ 400 is a compact and mobile machine, two people should move it over rough surfaces or up and
down grades.
2–2–11 Adjusting System Clock
Set the system clock for the LOGIQ 400 to the local time. For procedure of adjusting the system clock, refer to 4–3–3
Utility Menu, (A) TIME ADJUSTMENT, in Chapter 4, FUNCTIONAL CHECKS.

Table of Contents

Other manuals for GE LOGIQ 400

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 400 and is the answer not in the manual?

GE LOGIQ 400 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 400
CategoryMedical Equipment
LanguageEnglish

Related product manuals