GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–210
REV 9
6–2–84 Mother IC (FRU No. 616)
Time Required
1 Hour
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–119 on page 6–211.
1. Turn OFF the system.
2. Remove the Front Base Cover (FRU 305). Refer to 6–2–35 on page 6–131.
3. Remove the Side Cover Right (FRU 302). Refer to 6–2–32 on page 6–128.
4. Remove the Rear Cover (FRU 303). Refer to 6–2–33 on page 6–129.
5. Remove the Neck Assy (FRU 201). Refer to 6–2–26 on page 6–114.
6. Remove the Mother IC.