GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
TM
8–82 OPTIONS
REV 5
8–8–6 Installing Swivel Lock
1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction
46–015902).
2. Unscrew inside two hexagonal bolts fixing the left-side front caster to the frame as shown in
ILLUSTRATION 8–90.
HEXAGONAL BOLTS REMOVAL
ILLUSTRATION 8–90