Configuration
2026419-033E CIC Pro™ 6-71
Back up and restore the CIC Pro center configuration
NOTE
The backup and restore procedure must only be completed from a remote
computer.
NOTE
The backup portion of this section should be performed at regular intervals,
whenever changes are made to CIC Pro center configuration settings, and after
each preventative maintenance is performed. Regular backups ensure a complete
system restore if it should become necessary.
You can back up and restore the following:
Care unit defaults. See page 6-71.
Custom defaults. See page 6-75.
Back up care unit default configuration settings
NOTE
The real-time trend graph setting is not saved during backup. It must be restored
manually.
1. If you have not already logged onto Webmin, Log on to the Webmin service
interface remotely via the CARESCAPE network IX network on page 4-7.
2. Click Configuration > Backup/Restore > Unit Defaults.
3. Click Backup. The default files are temporarily written to the CIC Pro center and
are prepared for downloading to your local computer.