Field replaceable units (FRUs)
2026419-033E CIC Pro™ 9-15
CAUTION
NETWORK DEVICE TIME SYNCHRONIZATION — When
adding a new device (e.g., CIC Pro center) to the CARESCAPE
Network, the existing devices on the CARESCAPE Network will
synchronize to the new device’s time. To prevent potential time
synchronization issues, you should set the new device’s time to be as
close as possible to the time (within a few seconds) used by the
existing GE devices on the CARESCAPE Network.
Check the status of installed licenses on page 7-4.
Check access to all other care units on page 7-7.
Check the pressures unit-of-measure on page 7-8 (for Chinese applications
only).
Verify that the Remote Service agent is enabled. See Enable or disable the
Remote Service Agent on page 6-18.
Check the laser printer connectivity. See Check status of installed printers on
page 7-3.
Check the digital writer printer connectivity. See Check status of installed
printers on page 7-3.
Check MultiKM (Multimouse) operation on page 7-8.
Check full disclosure report printing on page 7-9.
Check read and write integrity of hard drive and solid state flash drive on
page 7-10.
Check audio component operation on page 7-16.
Check integrity of system files on page 7-20.
Check operation of the secondary display on page 7-4.
Check COMM ports on page 7-13.
Check BIOS information on page 7-13.
Check drive operation information (Flash drive and hard drive) on page 7-
14.
Check operation of the Watchdog countdown function on page 7-18.
Check video function and status of video card and drivers on page 7-19.
Check internal hardware temperature and voltage status on page 7-12.
Check system resources. See Environment Monitor messages on page 8-19.
Verify that the power LED is on after the unit is powered up.
NOTE
Do not re-image the Flash drive or hard drive after replacing the CPU PCB.
Replacing the CPU battery
Once the battery is replaced, the system loses its time and the time must be reset
before connecting the CIC Pro center to the existing CARESCAPE Network.