Field replaceable units (FRUs)
2026419-033E CIC Pro™ 9-19
13. Configure the system as necessary as described in Configuration on page 6-1.
NOTE
If the System Resource Indicator turns yellow or red during configuration,
refer to Environment Monitor messages on page 8-19.
14. Complete Configuration checkout procedures on page 7-3.
Replacing the SATA drive cables
CAUTION
NETWORK DEVICE TIME SYNCHRONIZATION — When
adding a new device (e.g., CIC Pro center) to the CARESCAPE
Network, the existing devices on the CARESCAPE Network will
synchronize to the new device’s time. To prevent potential time
synchronization issues, you should set the new device’s time to be as
close as possible to the time (within a few seconds) used by the
existing GE devices on the CARESCAPE Network.
1. Prepare the unit for disassembly on page 9-12.
2. Remove the cover on page 9-13.
3. Remove the two SATA cables that connect to the PCB CPU.
4. Remove the two SATA cables that connect to the hard and Flash drives.
5. Replace the SATA cables.
6. Re-assemble the cover on page 9-24.
7. Perform the following checkout procedures:
For the HDD SATA:
Perform Ground continuity test on page 10-6.
Perform Ground (earth) wire leakage current test on page 10-7.
Perform Enclosure leakage current test on page 10-8.
Check read and write integrity of hard drive and solid state flash drive on
page 7-10.
Check drive operation information (Flash drive and hard drive) on page 7-
14.
Check system resources. See Environment Monitor messages on page 8-19.
For the Flash drive SATA:
Perform Ground continuity test on page 10-6.
Perform Ground (earth) wire leakage current test on page 10-7.
Perform Enclosure leakage current test on page 10-8.
Check read and write integrity of hard drive and solid state flash drive on
page 7-10.
Check integrity of system files on page 7-20.
Check system resources. See Environment Monitor messages on page 8-19.