Upgrade software
2026419-033E CIC Pro™ 12-3
Prepare the CIC Pro center on page 12-3.
Prepare the service PC on page 12-4.
Install the software on the target CIC Pro centers on page 12-6.
Activate the software packages on page 12-8.
Complete the checkout procedures on page 12-9.
Establish alternate monitoring methods
When a care unit has two or more working CIC Pro centers, you can move monitored
patients from one CIC Pro center to another system if the CIC Pro center has empty,
unlocked patient slots available.
Gather the required equipment
The following equipment is required to upgrade CIC Pro center software:
Service PC (desktop or laptop computer) equipped with an Ethernet network
card, and running Windows NT, 2000, or XP. This PC functions as the software
transfer server.
Ethernet crossover cable.
This manual. See Configuration on page 6-1 and Complete configuration
checkout procedures on page 6-91.
Software upgrade CD for CIC Pro center, supplied as part of the software
upgrade kit.
Prepare the CIC Pro center
The CIC Pro center must be running and be connected to the CARESCAPE Network
IX network.
NOTE
The software installation process does not require the CIC Pro center to be
disconnected from the CARESCAPE Network IX or MC networks. Do not
stop the active monitoring of patients via the CIC Pro center. The software
upgrade files remain inactive on the target CIC Pro centers until you are
ready to activate them. See Install the software on the target CIC Pro centers
on page 12-6.
To activate the software package files on a target CIC Pro center, you must
use the local Webmin service interface tool to activate the software. The CIC
Pro center automatically reboots and begins to run the updated software.
Rebooting a CIC Pro center stops active patient monitoring. You will need to
plan for and establish alternate patient monitoring while you are activating
the software upgrade files on the target CIC Pro centers. See Activate the
software packages on page 12-8.