Equipment overview
2026419-033E CIC Pro™ 2-9
Touchscreen displays
The touchscreen display allows you to select any selectable screen object by gently
tapping the object with your finger.
The following guidelines apply when using a touchscreen display:
Applying tape or other items to the screen impairs the touchscreen’s
functionality.
Using pencils, pens, or other sharp, pointed objects can damage the touchscreen.
Displaying right click menus is not supported by the touchscreen.
Primary and secondary displays can be a combination of touchscreen and non-
touchscreen displays.
Mirror display
Networked remote displays can provide a duplicate (mirror image) view of a primary
CIC Pro center. When external speakers are connected to the remote displays, audible
alarm tones can also be sounded.
Laser printer
WARNING
SHOCK HAZARD — Laser printers are UL 60950/IEC 60950
certified equipment, which may not meet the leakage current
requirements of patient care equipment. This equipment must not be
located in the patient environment unless the medical system
standard IEC 60601-1-1 is followed.
Do not connect a laser printer to a multiple portable socket outlet
(MPSO) supplying patient care equipment. The use of an MPSO for
a system will result in an enclosure leakage current equal to the sum
of all the individual earth leakage currents of the system if there is
an interruption of the MPSO protective earth conductor.
The supported printers include:
HP LaserJet 2430 (USB and Network)
HP LaserJet P3005 (USB and Network)
HP LaserJet P3015 (USB and Network)
HP LaserJet 4000 (Network only)
HP LaserJet 4050 (Network only)
HP LaserJet 4100 (Network only)
HP LaserJet 4200 (Network only)
HP LaserJet 4250 (Network only)
A laser printer can be configured to print alarm graphs, control settings, waveforms,
parameter numeric data, events, full disclosure, and graphic trend data.