Configuration
4. Right-click the cic_xxxxxxxxx.cfd link and select Save Target As to save the
backup file.
Planning your care unit defaults restore strategy
NOTE
Understand the functionality as described below before you proceed with the
restore procedure.
NOTE
The real-time trend graph setting is not saved during backup. It must be restored
manually.
When planning a restore strategy in a care unit running more than one CIC Pro center,
it is important to know that when a CIC Pro center starts up, it will update some of its
configuration data to match any other CIC Pro centers it finds on the care unit
network. The exact data that is automatically updated in this fashion is shown in Data
module detail on page B-2.
In general, this would be data common to the care unit that should be consistent
between all the CIC Pro centers within that care unit. All other configuration data is
specific to the local CIC Pro center only, and is not shared between the CIC Pro
centers in the care unit.
See the following scenarios for information about how this automatic update impacts
the backup and restoration process of a CIC Pro center.
Scenario 1: Restore a CIC Pro center in a care unit running only one CIC Pro
center on page 6-72
Scenario 2: Add or upgrade a CIC Pro center in a care unit running multiple CIC
Pro centers on page 6-74
Scenario 3: Restore multiple CIC Pro centers in a care unit to a previous
configuration on page 6-74
Scenario 1: Restore a CIC Pro center in a care unit running only one CIC Pro
center
In this situation, automatic update does not occur, and is not a factor. All
configuration data written to the CIC Pro center during the restoration process
remains intact, since no other CIC Pro centers exist to trigger an automatic update.