7-16 CIC Pro™ 2026419-033E
Checkout procedures
Check audio component operation
NOTE
The CIC Pro center application must be stopped before conducting this test.
CAUTION
During shutdown or while in administrator mode, beds displayed by
the CIC Pro center will be unmonitored if not displayed by a
different CIC Pro center.
1. Log on to the CIC Pro center as Administrator on page 4-3.
2. From the desktop, launch Internet Explorer.
3. If you have not already logged onto Webmin, Log on to the Webmin service
interface on page 4-7.
4. Click Diagnostics > Preventative Maintenance > Audio Test.
5. On the Audio Test screen, select CIC Alarm Sounds from the drop-down list and
click Submit Query.