Service interfaces
2026419-033E CIC Pro™ 4-7
Webmin interface access
CAUTION
Users must log off of Webmin to prevent unauthorized access.
Closing Webmin is not equivalent to logging off. You must click
Logout.
You can access the Webmin service interface locally from the CIC Pro center, or
remotely from a service PC connected to the CARESCAPE Network IX network.
Log on to the Webmin service interface
Log on to the direct access (local) Webmin service interface
Complete the following procedure to log on to the Webmin service interface using the
CIC Pro center’s Browser function.
1. From the multi-patient viewer, click Browser.
2. Depending upon browser configuration, click on either the Favorites menu or the
Favorites button.
3. Click LocalWebmin.
4. In the Username field, type biomed.
5. In the Password field, type [password] and press Enter.
NOTE
The password is Change Me unless it has been changed.
See Change the logon password for Webmin on page 8-32.
Log on to the Webmin service interface remotely via the
CARESCAPE network IX network
1. Set up the service PC’s network properties. Refer to the table on page 4-8.