Preventive maintenance
2026419-033E CIC Pro™ 10-13
The system resource indicator should be checked periodically, and appropriate action
should be taken when needed.
Depending upon the color of the CIC Pro center system resource indicator, different
actions are required:
System resource restart procedure
When you are prompted to restart, always perform the safe shutdown procedure as
described.
1. From the multi-patient viewer, select Setup CIC > Service Password.
2. Type mms_com and then press Enter.
3. At the c:\Program Files\Marquette\CIC\<version#> prompt, type stop and press
Enter.
Yellow Warning! Available system resources are running low. A
Preventive Maintenance action is required.
Red Warning! Available system resources are running low.
System restart is required to correct the problem. Patients
will not be monitored at this Central while the System is
restarting. If the system is not restarted now, it will restart
automatically in approximately XX minutes.
NOTE
Where “XX” represents the number in minutes ranging
from 0 to 60.
Icon color Text
System resource indicator System resource condition Action required
Green Normal - System resources are within
normal parameters
No service action is currently required.
Yellow (value equal or greater than 364) Advisory - System has been running
continuously for at least 364 days without a
system resource reset. System resources
may be running low
Schedule a system resource restart.
Yellow (value less than 364) Advisory - System resources are running
low
Contact GE Service.
Red (value equal or greater than 365) Low - System has been running
continuously for at least 365 days without a
system resource reset. System resources
may be running low
Perform a system resource restart at the
earliest possible opportunity as part of
annual preventative maintenance.
Red (value less that 365) Low - Available system resources are
running critically low.
Contact GE Service.
Any color other than green (value less than
10)
Low - Available system resources are
running critically low.
Perform a system resource restart and
contact GE Service immediately. Be
prepared to provide log files. For more
information, refer to Access current log
files on page 8-21.