5-12 CIC Pro™ 2026419-033E
Installation
Connect the cables and peripheral devices
CAUTION
Do not insert the CARESCAPE Network IX or CARESCAPE
Network MC cable into the CIC Pro center at this time. Network
connectivity is enabled after the CIC Pro center has been installed
and configured for use.
Connect the keyboard and mouse
NOTE
Always position the mouse on a flat surface to prevent erratic mouse
movements and behaviors.
USB ports support passive extenders (per USB 2.0 standards) and does not
support active extenders. For more information, see Restrictions on page 5-
11.
Insert the keyboard and mouse cables into any available USB port on the CIC
Pro center.
Connect the external speakers
WARNING
ALARMS — Do NOT rely exclusively on the audible alarm system
for Bedside Monitoring. Adjustment of CIC Pro center alarm
volume to a low level or OFF during Bedside Monitoring may result
in inability to hear the alarm and a hazard to the patient. Remember
that the most reliable method of Bedside Monitoring combines close
personal surveillance with correct operation of monitoring
equipment.
After connecting the monitor to the central station and/or nurse-
alert, verify the function of the alarm system. Repeat this
verification periodically, including a check of all connected
speakers.
CIC Pro center audible alarms will not sound for patients with
bedside monitoring devices configured to “Operating Room” mode.
The functions of the alarm system for monitoring of the patient must
be verified at regular intervals. Check speaker volume periodically
to ensure audio alarm functionality.
In addition to two internal speakers, the CIC Pro center is equipped with a 1/8-inch
external speaker connection port.