9-18 CIC Pro™ 2026419-033E
Field replaceable units (FRUs)
Check drive operation information (Flash drive and hard drive) on page 7-
14.
Check system resources. See Environment Monitor messages on page 8-19.
Verify that the power LED is on after the unit is powered up.
Replacing the flash drive
CAUTION
NETWORK DEVICE TIME SYNCHRONIZATION — When
adding a new device (e.g., CIC Pro center) to the CARESCAPE
Network, the existing devices on the CARESCAPE Network will
synchronize to the new device’s time. To prevent potential time
synchronization issues, you should set the new device’s time to be as
close as possible to the time (within a few seconds) used by the
existing GE devices on the CARESCAPE Network.
1. Prepare the unit for disassembly on page 9-12.
2. Remove the cover on page 9-13.
3. Remove the power cable from the hard drive.
4. Remove the SATA cable from the hard drive.
5. Remove the power cable from the Flash drive.
6. Remove the SATA cable from the Flash drive.
7. Loosen the four screws that connect the Flash drive to both sides of the mounting
bracket.
NOTE
The mounting bracket holds both the hard drive and the Flash drive. The
hard drive is on the top; the Flash drive on the bottom.
8. Slide the Flash drive out of the mounting bracket.
9. Replace the Flash drive.
10. Reverse steps to re-assemble.
11. Re-assemble the cover on page 9-24.
12. Re-image the Flash drive. See option 1 on page 11-5.