Upgrade software
2026419-033E CIC Pro™ 12-9
CAUTION
If a USB memory stick other than the stick containing the reload
image is plugged into the CIC Pro center, you need to remove it
prior to rebooting. Otherwise, the message Operating system
missing will be displayed.
Complete the checkout procedures
NOTE
After upgrading the CIC Pro center software from v5.0.x to CIC v5.1.x, the
existing Multimouse configurations are lost. Before proceeding to checkout, you
must re-configure Multimouse on the CIC Pro center and ensure that the
Multimouse functionality regained. See Perform MultiKM (Multimouse) setup
on page 6-77.
NOTE
The Log Compression Configuration Utility must be run:
After upgrading to CIC Pro center software v5.1.x
When a CIC Pro center v5.1.x is added to an existing Multimouse group
Initial configuration of any CIC Pro center v5.1 device during new installs
During CIC Pro center v5.1 configuration after software re-image/ reload
process
After activating the installed software and before using the CIC Pro center to monitor
patients, you must verify proper operation of this device in the patient care and
networking environments.
To verify proper operation in the patient care and networking environments, complete
all of the checkout procedures identified in Complete configuration checkout
procedures on page 6-91.