Configuration
2026419-033E CIC Pro™ 6-89
NOTE
Depending on the original status of the USB memory stick, at 99% on the
progress indicator, the reload process may appear to be idle. The flashing
LED on the USB memory stick indicates that the program is still processing.
If the USB memory stick is not equipped with an LED, wait for
approximately five minutes after the progress indicator shows that the
process is 99% complete to proceed.
8. Remove the USB memory stick from the service PC.
NOTE
The Make USB stick process also prepares the USB memory stick for safe
removal from the service PC. It is not necessary to take additional steps to
safely eject the USB memory stick.
9. Close the GE Patch Distribution Utility window.
10. Remove the Log Compression Configuration Utility software CD from the
service PC.
Run Log Compression Configuration Utility
WARNING
LOSS OF MONITORING — This procedure requires you to stop
the CIC Pro center application and reboot the CIC Pro center after
completion. Stopping the application or rebooting a CIC Pro center
stops active patient monitoring. Notify the site’s biomedical
department to provide alternate bedside patient monitoring. See
Establish alternate monitoring methods on page 12-3.
1. Insert the USB memory stick containing the Log Compression Configuration
Utility into one of the available USB ports of the CIC Pro center.
2. From the multi-patient viewer, click Setup CIC.
3. Click the Service Password tab.
4. Type mms_com as the password and press Enter.
5. At the Windows command prompt, type stop and press Enter.
6. Type su administrator admin1,3,5,7 and press Enter.
7. In the new command window, type H: and press Enter.
8. At the H:\ prompt, type LogCompressionConfigUtility_Install.bat and press
Enter.
9. Once started, the message This V5.1.0 CIC has not been reconfigured;
Continue with reconfiguration of this system [y/n]? will be displayed. Type y
and press Enter to continue.