Checkout procedures
2026419-033E CIC Pro™ 7-7
Check operation of audible alarm tones
WARNING
The CIC Pro center audible alarms will not sound for bedside
patients configured to the Operating Room mode.
NOTE
All clinical setting values must be received from the GE Clinical Application
Specialist (CAS) or from the nursing director of the care unit. Consult with your
CAS or hospital staff about the use of these functions and if they should be used
with your configuration.
NOTE
Active alarms override this test.
1. From the single patient viewer menu, click Monitor Setup > Alarm Control.
2. Click Alarm Help > Advisory Alarm.
3. Verify the alarm sounds through both speakers.
Check access to all other care units
1. From the multi-patient viewer menu, click View Other.
2. Verify that the list shows all care units that should be networked.
Check the operation of the Citrix application viewed at the CIC Pro
center
It is the responsibility of the hospital’s IT or Biomedical department to meet the
required CIC Pro center configuration needs to check this functionality.
Check the hospital intranet browser functionality
It is the responsibility of the hospital’s IT or Biomedical department to meet the
required CIC Pro center configuration needs to check this functionality.
Check the status of locked beds
Based on the needs of the hospital staff, ensure that the required beds are locked. See
Set up locked beds on page 6-53.
Check the language settings
In the multi-patient viewer window, check if the patient names are displayed the
appropriate language.