9-16 CIC Pro™ 2026419-033E
Field replaceable units (FRUs)
WARNING
DATA LOSS—It is required that you set the time as close as
possible to any other device on the CARESCAPE Network before
connecting the CIC Pro center to the existing CARESCAPE
Network. Otherwise, abrupt data loss may occur and CIC Pro center
performance may be corrupted.
NOTE
This part is not stocked by GE service and can be purchased from your local
store. When purchasing from a retail supplier, use the specification number on
the battery to ensure an equivalent battery is purchased.
1. Prepare the unit for disassembly on page 9-12.
2. Remove the cover on page 9-13.
3. Push the battery latch away from the battery until the battery pops loose and
remove the battery.
4. Replace the battery.
5. Reverse steps to re-assemble.
6. Re-assemble the cover on page 9-24.
7. Set the time on the CIC Pro center as close as possible to any other device on the
CARESCAPE Network.
CAUTION
NETWORK DEVICE TIME SYNCHRONIZATION — When
adding a new device (e.g., CIC Pro center) to the CARESCAPE
Network, the existing devices on the CARESCAPE Network will
synchronize to the new device’s time. To prevent potential time
synchronization issues, you should set the new device’s time to be as
close as possible to the time (within a few seconds) used by the
existing GE devices on the CARESCAPE Network.
8. Perform the following checkout procedures:
Perform Ground continuity test on page 10-6.
Perform Ground (earth) wire leakage current test on page 10-7.
Perform Enclosure leakage current test on page 10-8.
Check access to all other care units on page 7-7.
Verify that the power LED is on after the unit is powered up.