REMOTE INTERFACE
329
Sweep mode is currently enabled.
SOURce[1|2|3|3RF]:FREQuency:STARt
Sets the start frequency of the sweep. 100Hz is the
default start frequency.
To sweep up or down, set the stop frequency
higher or lower than the start frequency.
SOURce[1|2|3|3RF]:FREQuency:STARt
{<frequency>|MINimum|MAXimum}
SOUR1:FREQ:STAR +2.0000E+03
Sets the start frequency to 2kHz.
SOURce[1|2|3|3RF]:FREQuency:STARt? [MINimum|
MAXimum]
Returns the start frequency
in Hz.
SOUR1:FREQ:STAR?
+2.0000000000000E+03
Returns the maximum start frequency allowed.
SOURce[1|2|3|3RF]:FREQuency:STOP
Sets the stop frequency of the sweep. 1 kHz is the
default start frequency.
To sweep up or down, set the stop frequency
higher or lower than the start frequency.
SOURce[1|2|3|3RF]:FREQuency:STOP
{<frequency>|MINimum|MAXimum}