REMOTE INTERFACE
333
SOURce[1|2|3|3RF]:SWEep:SOURce
Source Specific
Command
Description
Sets or queries the trigger source as immediate
(internal), external or manual. Immediate (internal)
is the default trigger source. IMMediate will
constantly output a swept waveform. EXTernal
will output a swept waveform after each external
trigger pulse. Manual will ouput a swept
waveform after the trigger softkey is pressed.
Note
If the APPLy command was used to create the
waveform shape, the source is automatically set to
IMMediate.
The *OPC/*OPC? command/query can be used to
signal the end of the sweep.
If the trigger source is set to manual, the function
generator starts sweeping each time a trigger
command is received. To trigger the function
generate from remote interface, it is necessary to
send a * TRG trigger command.
Syntax
SOURce[1|2|3|3RF]: SWEep:SOURce
{IMMediate|EXTernal| MANual}
Example
SOUR1: SWE:SOUR INT
Sets the sweep source to internal.
Query Syntax
SOURce[1|2|3|3RF]: SWEep:SOURce?
Return Parameter
IMM
Immediate
EXT
External
MANual
Manual
Example
SOUR1:SWE:SOUR?
IMM
The sweep source is set to internal.

Table of Contents

Other manuals for GW Instek MFG-2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek MFG-2000 Series and is the answer not in the manual?

GW Instek MFG-2000 Series Specifications

General IconGeneral
BrandGW Instek
ModelMFG-2000 Series
CategoryInverter
LanguageEnglish

Related product manuals