Roche Diagnostics
Operatorโs Manual ยท Version 3.4 B-49
Cedex Bio analyzer 5 Daily operation
Analyzing samples
Monitoring the analysis progress
During operation, you should regularly check the following items:
o The status of the buttons on the Overview tab
o The Alarm Monitor
Checking the sample tube status
a To check the sample status
1
Press the Overview tab.
Interpreting the sample tube
buttons
A Sample tube buttons
B Information on tests that are associated
with defined orders
C Information on results that are associated
with defined orders
D Alarm Monitor
Figure B-12
Icon Meaning Possible action
1
The number in the button denotes
the position on the sample area.
n/a
A sample tube button with a wide
edge symbolizes a STAT order.
n/a
All tests are accepted. You can remove the sample tube.
All tests are pipetted. You can remove the sample tube, the
time to completion disappears when
you do so.
All tests are completed but not yet
accepted.
Validate the results.