Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.4 F-33
Cedex Bio analyzer 14 ISE operation
Replacing ISE fluid bottles
Replacing ISE fluid bottles
ISE fluid bottles are supplied with a barcode that contains information on their
expiration date, which is monitored by the system.
Tools and materials required o ISE Calibrator indirect or ISE Reference Solution bottle
o Tissues
a To replace an on-board ISE fluid bottle
1
Check that the system is in Standby status.
2
Choose Overview > .
3
Check the status of the ISE bottles (A).
Roche recommends replacing ISE fluid bottles as soon as their expiration date has passed.
(Their icon on the screen turns orange.)
For information on the stability of fluids, see their Instructions for Use.
WARNING
Injury through working solutions
Direct contact with cleaning solutions or other working solutions may cause personal
injury.
When handling such solutions, exercise the precautions required for handling them,
observe the cautions given in the Instructions for Use, and observe the information given in
the Safety Data Sheets available for Roche Diagnostics cleaning solutions.