Roche Diagnostics
Operatorโs Manual ยท Version 3.4 A-51
Cedex Bio analyzer 3Hardware
Hardware overview
External waste container The yellow external waste container is placed on the external fluid rack. It is designed
to be washed and reused.
Because the system periodically performs wash actions, an external waste container
must be connected at all times. Therefore, when you empty the waste container, you
immediately replace it with the spare container and then empty the original
container. (The analyzer is supplied with a spare waste container.)
There is no active level monitoring for the external waste container, but you are
notified if the external waste container has not been emptied for more than one day.
The external waste container is connected by tubing to the internal waste tank.
e
See Connecting and disconnecting the external fluid containers on page B-137.
See Checking the status of the external fluid containers on page B-17.
See maintenance action Clean the water and waste containers on page D-16.
Water container
The white water container is positioned on the external fluid rack. Attached to the
cap is a suction tube, which is equipped with a water filter.
There is no active level monitoring for the water container, but you are notified if the
water container has not been refilled for more than one day.
The water container is designed to be washed and refilled.
The water container is connected by tubing to:
o Wash pump
o Syringe assembly
o Probe
e
See Connecting and disconnecting the external fluid containers on page B-137.
See Checking the status of the external fluid containers on page B-17.
See maintenance action
Clean the water and waste containers on page D-16.
See maintenance action Replace water inlet filter on page D-18.
A Tubing
B Container
C Water inlet filter
Figure A-19 Water container