Page #324 background imageLoading...
Page #324 background image
Roche Diagnostics
B-184
Operatorโ€™s Manual ยท Version 3.4
7 Configuration Cedex Bio analyzer
Configuration
Short guide to extra wash cycle definitions
f Utilities > Applications > Extra Wash Cycles >
Extra wash cycle definitions
Item Values Comment
Active On, Off Defines whether the extra wash cycle can be applied
or not.
Action None
Before
After
Defines when, in relation to the pipetting action from
the bottle defined in Trigger Bottle Code and Pipet-
ting Type, the extra wash cycle should be performed.
Trigger-Bottle
Code
0 ... 999 Bottle set code of the fluid that triggers performance
of the extra wash cycle.
If you use the pipetting type Sample, it defines the
bottle whose fluid will be prevented from being car-
ried over to the sample by using the extra wash cycle.
Pipetting Type Reagent 1
Reagent 2
Start Reagent
Sample
Defines the individual bottle (bottle of the set or a
sample tube) whose pipetting triggers performance of
the extra wash cycle.
Cleaner-Bottle
Code
0 ... 999 Material code of the cleaner.
Volume ฮผL Amount of cleaner that will be pipetted during the
extra wash cycle.
Table B-34 Extra wash cycle definitions