Roche Diagnostics
Operatorโs Manual ยท Version 3.4 B-115
Cedex Bio analyzer 6 Special operations
Exporting data
Interpreting result data
The following tables list, for the sample, calibration, and QC result files, the column
headings of the spreadsheet and provide information on the kind of information the
columns contain.
res_yyyymmddhhmmss.csv
Column Title Description
Instr Instrument ID (always 30 for Cedex Bio)
Msg Message Type:
7 Result for sample, control, or calibration
Index 166 ID of the data set used
App Application code
Date Date when the result was calculated
Time Time when the result was calculated
SW-Version Software version installed on the analyzer
Serial Serial number of the analyzer
Te st Short name of the application
User User name ($SYS$ if Auto Accept is on)
Sample Sample ID
Order-Time Time when the order was started
Result Results of the measurement
Unit Defined display unit
Flags Flag name
Rates Rate value calculated in Abs. (Last Calculation Point - First
Calculation Point)
Raw1 - Raw40 Absorbance value of cycle 1 through 40 in Abs. (one cycle = 18 s)
isSTAT Generated as part of a STAT order
Table 33 Explanations on test results