Cedex Bio analyzer 11 Result flags
List of flags
? Cal
Meaning A result where its calibration is flagged.
Message ID 110
Priority 29
Possible cause The calibration was in question and has been flagged, however the calibration could
be used to calculate the result (compare with Cal Error where the calibration did not
provide a usable result).
Recommended actions Check and redo calibration if needed.
Cal Error
Meaning No valid calibration data available.
Message ID 43
Priority 24
Possible cause Caused by an alarm that occurred during calibration because the calibration could
not provide a usable result.
Recommended actions Check the flags of the calibration and proceed to deal with the calibration first.
Calc Error
Meaning Calculation error.
o General calculation error.
o Slope or nonlinear standard curve cannot be calculated due to a calibration error.
Message ID 16
Priority 13
Possible cause o Calibrator outdated or deteriorated.
o Misplacement of the calibrator tubes.
Recommended actions o Repeat calibration with fresh calibrators.
o If the calibrators were incorrectly positioned, replace the calibrators and repeat
the calibration.