Roche Diagnostics
Operatorโs Manual ยท Version 3.4 B-179
Cedex Bio analyzer 7 Configuration
Configuration
Result handling
Calibration
QC
Item Values Comment
Sample Auto
Accept
On, Off Results that do not contain a flag are automatically
accepted. Results with flags that are marked in a list of
acceptable flags are accepted as well.
Sample Accept.
Flags
List of possible flags. Results with marked flags are
automatically accepted if Sample Auto Accept is on.
QC Auto Accept On, Off Results that do not contain a flag are automatically
accepted. Results with flags that are marked in a list of
acceptable flags are accepted as well.
QC Acceptable
Flags
List of possible flags. Results with marked flags are
automatically accepted if QC Auto Accept is on.
Cal Auto Accept On, Off Results that do not contain a flag are automatically
accepted. Results with flags that are marked in a list of
acceptable flags are accepted as well.
Cal Acceptable
Flags
List of possible flags. Results with marked flags are
automatically accepted if Cal Auto Accept is on.
Table B-27 Result handling definitions
Item Values Comment
Forecast Hours 0 ... 999 h Calibrations that fall due within this period will be
performed collectively.
0 means working without this feature.
The default value is 8.
Table B-28 Calibration definitions
Item Values Comment
Rule 1 Rule: A flag is generated if one result is outside a
limit.
s : Standard deviation. The value for the standard
deviation is defined in the QC lot.
Off
2.5s
3s
Rule 2 Rule: A flag is generated if two consecutive results are
outside the same limit.
s : Standard deviation. The value for the standard
deviation is defined in the QC lot.
Off
2s
Table B-29 QC definitions