Cedex Bio analyzer 6 Special operations
Importing data
Importing certificates
Certificates are digital records that ensure the authenticity of reagent barcodes. On
each Cedex Bio analyzer there must be a certificate installed.
a To import certificates
1
Choose Utilities > Import > Certificate.
A screen is displayed, that lists the currently installed certificates.
2
Press .
A screen is displayed, asking you to insert the USB stick.
3
Insert the USB stick.
4
Press .
5
Select the certificate.
You recognize a certificate by its file extension .prm.
6
Press to confirm the selection.
The certificates are installed on the system.
The screen is displayed again, that lists the currently installed certificates.
Importing automatic mixing and extra wash cycle definitions
Mixing and extra wash cycle information is contained in a separate barcode on a
barcode transfer sheet.
a To import mixing and extra wash cycle information
1
Choose Utilities > Import > EWC/Mixing.
A screen is displayed, asking you to scan the barcode.
2
Scan the barcode.
The data are installed on the system. A message will inform you when the
installation is complete. The screen for importing data is displayed again.
e
For information on defining extra wash cycles see Defining extra wash cycles on
page B-182.
After the database installation
o Load all bottle sets that are defined, using the appropriate software features.
o Perform Abs. Air/Water Calibration maintenance action.
o If TRL Check is activated, the value for the lower limit of the measuring range is not
overwritten when importing the database. Check the relevant application documenta-
tion for the exact value, you may need to adjust it manually. See Calculation definitions
on page B-152. (Note that there may be applications where the lower limit of the mea-
suring range cannot be changed.)