Roche Diagnostics
Operatorโs Manual ยท Version 3.4 B-55
Cedex Bio analyzer 5 Daily operation
Analyzing samples
Stopping and restarting a run
a To stop a run
1
Press the global action button.
A screen is displayed that offers several kinds of stopping, each of them
representing a certain level of interrupt.
Interpreting the screen
Figure B-15
Abort When the system is in Operating status: Stop immediately all pro-
cessing activities.
Pipettings that were not finished are considered not pipetted.
Measurements that have not yielded a valid result are considered not
measured.
Recover When in Standby or Stopped status: Initialize all systems and mod-
ules that are currently not ready.
Sampling stop Finish the current pipetting action, but do not start a new one.
You can restart processing by pressing the
global action but-
ton
.
Stop Printing Stop the current printing task. (It may take a few moments before
printing actually stops.)
Shutdown Shut down the Cedex Bio software and the operating system.
This option can only be performed in Standby status.
Restart Shut down the Cedex Bio software and automatically restart it.
This option is available in Standby status only. It is used in cases
where a configuration change requires restarting the software for it
to become effective.