12 Troubleshooting Cedex Bio analyzer
Detailed procedures
Realigning the reagent disk
a To realign the reagent disk
1
Insert your fingers in the holes in the reagent disk cover an lift the reagent disk.
2
Insert a pencil in the opening in the center of the underside of the reagent disk.
3
Turn the reagent disk until the arrows of the reagent disk and the cover face each
other.
Spillage through tipping reagent disk
When handling the reagent disk, make sure not to tilt it.
A Incorrect position: The two arrows do not
face each other.
B Correct position: The two arrows face each
other.
Figure E-7