Roche Diagnostics
F-54
Operatorโ€™s Manual ยท Version 3.4
15 ISE maintenance Cedex Bio analyzer
ISE maintenance actions
Clean ISE tower automatically
f Utilities > Maintenance > Clean ISE Tower Automatically
The ISE tower must be cleaned regularly to ensure the proper functioning of the ISE
unit. Cleaning includes deproteinizing and activating the ISE tower.
Operator time Approximately 3 minutes.
Prerequisites The system must be in Standby status.
Tools and materials required m Activator
m ISE Deproteinizer
a To clean the ISE tower automatically
1
Select the maintenance action Clean ISE Tower Automatically.
2
Press .
The maintenance definition screen is displayed.
3
Press to start the action.
A placement list is displayed.
4
Place the Activator and the ISE Deproteinizer on the sample area positions
indicated on the screen.
5
Press to confirm the placement and to start the action.
A message will inform you when the maintenance action is complete.
6
Remove the Activator and ISE Deproteinizer from the sample area.
7
Press .
Make sure that you have read and understood section Safety information on page F-46.
The following warning messages in particular are relevant:
o Injury through working solutions on page F-46.
o Infection by biohazardous materials on page F-46.