Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.4 E-9
Cedex Bio analyzer 10 Messages and alarms
Alarm monitor
Confirming alarm messages
Confirming a message means that you have dealt with the issue mentioned in it.
a To confirm a message
1
Press .
2
Select the message. (Use and to display the previous and next message
respectively.)
3
Press .
A confirmation dialog box is displayed.
o Press All to remove all messages from the list.
o Press Selected to remove the currently displayed message.
The message or messages are removed from the Not confirmed list, but they
remain in the other lists and the alarm messages log, which you can export.
When all messages are confirmed the Alarm button is turned off (not lit).
Exporting the alarm message log
a To export the message log file
1
Press .
2
Press .
A screen is displayed asking you to insert a USB stick.
3
Insert the USB stick.
4
Press to confirm the insertion.
A screen is displayed for selecting a directory.
Press *.tgz to display log files only.
Press *.* to display all files and directories.
5
Select a directory for the error log.
6
Press to confirm the selection.
All log files are added to a file with the name format log_yyyymmddhhmmss.tgz
and copied to the stick. The error log file has the name format
err_yyyymmddhhmmss.log.