Roche Diagnostics
Operatorโs Manual ยท Version 3.4 B-103
Cedex Bio analyzer 6 Special operations
Calibration
Calibration
Deleting calibration results
Obsolete calibration results are automatically deleted as part of the daily end of shift
activities if there are more than five obsolete calibration results for the test.
(Calibration results become obsolete when the empty set is removed or if it was
removed more than 30 days ago.)
This section describes how to proceed if you need to delete results outside the normal
daily routine.
e
For deleting calibration results at the end of a shift, see Cleaning up the database on
page B-92.
a To delete a calibration
1
Choose Workp lac e > Calibrations.
2
Select the calibration entry.
3
Press .
A confirmation screen is displayed.
4
Press .
The order and the results are deleted. You cannot retrieve them any more from
the system.