Roche Diagnostics
A-30 Operatorโs Manual ยท Version 3.4
2 Introduction to the analyzer Cedex Bio analyzer
Daily operation
Calibration status of a set
Each reagent set has one of the following calibration statuses:
CU (current) denotes that the set is on board and that its calibration results are currently
used.
OB (obsolete) denotes that the setโs calibration results are no longer used.
This status applies for example to the following situations:
o The set was removed and it is empty.
o The set was removed and it is not empty. It was removed more than 30 days ago.
SB (standby) indicates that the setโs calibration results are not currently used.
This status applies for example to the following situations:
o A new set was loaded and calibrated while an identical set was still in use (pre-
calibration).
o The set was removed not more than 30 days ago and it is not empty.
Calibration result storage
The current and up to five obsolete calibration results are stored on the system. If
there are more than five obsolete calibration results, the oldest obsolete calibration
results are automatically deleted as part of the daily end of shift activities.
Validating calibration results
Applications define checks for ranges and limits. If these are exceeded, the results are
flagged.
Each new calibration result has to be validated. If flags were generated, you must
determine their cause and decide whether to accept the result, rerun the calibration
or continue using the old calibration results.
You can automatically accept unflagged results and results with flags that are
contained in a specific list of flags that should be ignored.
e
See Editing the acceptable flags list on page B-174.
Calibration procedures
There are three basic procedures for performing calibration:
o Calibrating all reagent sets that need calibrating
o Calibrating all reagent sets that will need calibrating during the forecast period
o Calibrating individual reagent sets
Forecast period The forecast period is a configurable period of time. Calibrations that fall due within
this period will be performed collectively.
e
See Calibration on page B-179.
Typically you would set this period to fit your shift length, for example 8 hours. This
would enable you to prepare the analyzer before the work shift starts and so avoid
having to interrupt sample processing for performing calibrations.