EasyManuals Logo
Home>Roche>Measuring Instruments>Cedex Bio

Roche Cedex Bio User Manual

Roche Cedex Bio
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #197 background imageLoading...
Page #197 background image
Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.4 B-57
Cedex Bio analyzer 5 Daily operation
Validating sample results
Validating sample results
The Cedex Bio analyzer provides several aids for validating results:
o In the results list, results that fall outside predefined technical ranges are flagged.
o You can display detailed result information that allows you to make a considered
decision.
o Non flagged results can be accepted automatically.
o You c an pr int t he re su lts .
o You can export the results and process them on an external computer.
The following ways of dealing with results are available:
o Accept the result.
o Re-perform the identical test (Repeat).
o Re-perform the test using a predefined different dilution (Rerun).
The effect of flagged results depends on the configuration (Utilities > Configuration
> Result Handling). The following table shows how.
e
See Editing the acceptable flags list on page B-174.
There are two main approaches to validating results:
o To validate sample results of an order whose sample is still on board, press the
tube button on the Overview tab.
o In all other situations, choose Work pl ac e > Result Review.
Results must be accepted before they can be transmitted to a host or before they can be
automatically printed.
Ratio results cannot be manually accepted. They are automatically accepted if their con-
stituent results are accepted.
Configuration setting Effect
Sample Auto Accept: On Results that do not contain a flag are auto-
matically accepted. Results with flags that are
marked in a predefined list of flags that
should be ignored are accepted as well.
Sample Auto Accept: Off All results need to be manually accepted.
Sample Accept Flags Flags that are marked in this list will be
ignored by the system.
Table B-4 Sample flag configuration and its effect

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio and is the answer not in the manual?

Roche Cedex Bio Specifications

General IconGeneral
BrandRoche
ModelCedex Bio
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals