EasyManuals Logo
Home>Roche>Measuring Instruments>Cedex Bio

Roche Cedex Bio User Manual

Roche Cedex Bio
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #525 background imageLoading...
Page #525 background image
Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.4 F-75
Cedex Bio analyzer 16 ISE troubleshooting
List of ISE flags
List of ISE flags
This section lists the general and ISE-specific flags in alphabetic order and provides
information on each of the flags.
e
For information on flags that concern absorbance measurements only, see List of flags on
page E-28.
>
Meaning Result generated by re-performing the test with the same running parameters
(repeated).
Message ID 120
Priority 32
Possible cause The user initiated the repeat.
Recommended actions No action required.
Air Fluid
Meaning o The length of the sample segment was too short.
o The end of the sample segment was not detected. It takes too long for a sample
segment to reach the electrodes.
Message ID 19
Priority 4
Possible cause o Air bubbles are present.
o The ISE tower is clogged.
o The dosing from the sample probes is inaccurate.
o The ISE tubing is leaking, blocked, or not conditioned.
o Mixing is too vigorous.
Recommended actions o Check the quality of the sample.
o Check the sample tubes.
o Check the ISE tower and ISE tubing for leaks and blockages.
o Repeat the measurement.
Where a flag description gives a list of recommended actions, work through each step in
turn until you have solved the problem. Proceed to the next step only after you have com-
pleted the previous one.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio and is the answer not in the manual?

Roche Cedex Bio Specifications

General IconGeneral
BrandRoche
ModelCedex Bio
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals