Roche Diagnostics
E-28
Operatorโ€™s Manual ยท Version 3.4
11 Result flags Cedex Bio analyzer
List of flags
List of flags
This section lists the general flags and those that concern absorbance measurements
only, and it provides information on each of the flags.
e
For information on ISE-specific flags, see List of ISE flags on page F-75.
>
Meaning Result generated by re-performing the test with the same running parameters
(repeated).
Message ID 120
Priority 32
Possible cause The user initiated the repeat.
Recommended actions No action required.
V
Meaning A rerun result with diluted sample.
Message ID 121
Priority 33
Possible cause The user initiated the rerun.
Recommended actions No action required.
Ag Excess
Meaning Antigen excess.
The sample contains an excess antigen and a valid result cannot be calculated.
Message ID 6
Priority 19
Recommended actions Rerun the test with postdilution.