Installation
2026419-033E CIC Pro™ 5-7
Un-interruptible power supply (UPS)
WARNING
Connect the UPS to the CIC Pro center(s) and display monitor(s)
only. Do not connect printers or other devices to a UPS, as such
devices may shorten estimated run-times. If AC line power is not
restored before UPS run-time is exceeded, the CIC Pro center
improperly shuts down and patients will not be monitored.
WARNING
Without a UPS, power line outages may result in:
Improper shutdown of the CIC Pro center, causing lengthy disk
scan procedures on reboot.
Data loss.
If power to the CIC Pro center is lost, patient monitoring
information will no longer be displayed or stored.
GE recommends using a UPS with the CIC Pro center. See Accessories on page 9-6
for a listing of UPS units available for the CIC Pro center. Follow the manufacturer’s
recommendations for installing the UPS.
NOTE
The CIC Pro center draws approximately 40 watts of power. The displays each
draw approximately 50 watts of power. See the table on page 5-7.
The following table identifies the estimated wattage for CIC Pro center MP100
systems and 19-inch medical displays:
The actual UPS run-times are variable, and are affected by the following conditions:
Battery age
Ambient temperature
Site-specific UPS usage patterns
Load characteristics
Number of CIC
Pro center
MP100
systems
Number of CIC Pro center
19-inch medical displays
Total load with CIC Pro center MP100
system(s) and CIC Pro center 19-inch
medical display(s) (estimated watts)
One One display (primary) 75 W
One Two displays (dual/
secondary)
115 W
Two Four displays (Two primary,
Two secondary)
230 W